General Technologies CM1000 : 1000 Amps AC/DC Current Clamp Meter User Manual

Cm1000, Safety rules, International symbols

Advertising
background image

1. SAFETY RULES

This meter is designed for indoor use at temperatures between 0°C to 40°C and

altitudes up to 2,000m.

To ensure that the meter is used safely, follow all safety and operating instructions

in this operation manual. If the meter is not used as described in this operation

manual, the safety features of this meter might be impaired.

Do not use the meter if the meter or test leads look damaged ,or if you suspect that

the meter is not operating properly.

When using the instrument, keep your fingers behind the finger guards on the

plastic casing and probes.

Disconnect the live test lead before disconnecting the common test lead.

Make sure power is off before cutting, desoldering, or breaking the circuit wires.

Small amounts of current can be dangerous.

Do not apply more than 1000 VDC or 1000V AC rms between a terminal and

ground.

To avoid electrical shock, use CAUTION when working above 60V DC or 25V AC

rms. Such voltages pose a shock hazard.

Never make measurements with the battery cover off.
To avoid electrical shock or damage to the meter, do not exceed the input limits.

2. INTERNATIONAL SYMBOLS

Important information

Dangerous Voltages

see manual

Continuity

AC

Ground

DC

Double Insulation

3. TECHNICAL SPECIFICATIONS

3.1 General Specifications

Display: 4 digits LCD, 9999 counts and engineering

units annunciator display

Polarity: Automatic, (-) negative polarity indication

Zero adjustment: One touch button

Sample rate: 0.5 Sec.

Peak hold sample time: 10 mS for DCV and DCA

Over range indication: Only the “OL” is displayed

Power: 2x UM-4 or AAA 1.5 Volt battery

Auto power off: Automatically powers off after 30 min. of

no operation

Battery life: Approx. 50 hours. (w/ alkaline batteries)

Clamp opening size: 25 mm maximum.

Dimension: 189 x 70 x 34 mm.

Weight: Approx. 220 g. (Including battery)

Accesories: User’s Manual, Soft Carrying Pouch, Test

Leads and 2 x AAA Alkaline Batteries

3.2 Electrical Specifications

The mark next to the probe tip, is a warning that the input voltage

should not exceed the indicated values. This is to prevent damage to

the internal circuitry.

The function switches should be set to the function which you want to

test before operation.

Accuracies are ±(% of reading + number of least significant digits) at

23°C ±5°C, less than 75% RH.

4. OPERATION

4.1 INSTRUMENT DESCRIPTION

1) Current sensing clamp

2) Safety protection ring

3) Clamp opening handle

4) LCD Display

5) COM input terminal

6) Positive input terminal

7) Zero button

8) Function selection knob

9) Peak/Data hold button

Function

Range

Resolution

Accuracy

Remarks

Overload Protection

DC Current

1000 A

0.1 A

±(2.5%+10)

-

1020 Arms

AC Current

1000 A

0.1 A

±(2%+10)

50~500Hz

1020 Arms

DC Voltage

1000 V

0.1 V

±(1%+2)

-

1020 Vrms

AC Voltage

1000 V

0.1 V

±(1.5%+5)

50~500Hz

1020 Vrms

Resistance

100K�

10 �

±(1%+3)

-

660 Vrms

Continuity

Beeps when < 1K�

3VDC Max. Test Voltage

660 Vrms

4.2 LCD Screen Description

4.3 Current Measurement

CAUTION: Remove all test leads connected to the instru-

ment before proceeding with this test.

4.3.1 DC Current Measurement
Set the function selector knob to (DCA) and press the Zero button,

the will appear in the display. Open the clamp by pressing the clamp

opening handle, insert the cable or wire inside the current sensing clamp,

and then close the clamp and read from the LCD screen.

4.3.2 AC Current Measurement
Set the function selector knob to (ACA). Open the clamp by pressing the

clamp opening handle, insert the cable or wire inside the current sensing

clamp, and then close the clamp and read from the LCD screen.

600V CATII

300V CATIII 600A

H

HOLD

OFF

A

A

V

V

ZERO

P H

A

600V CATII

300V CATIII

600V

COM

1

2

3

4

5

8

9

7

6

P

PEAK

Press

2 Sec

General Technologies Corp.

CM1000

1000 Amp AC/DC Clamp Meter

600V CATII

300V CATIII 600A

H

HOLD

OFF

A

A

V

V

ZERO

P H

A

600V CATII

300V CATIII

600V

���

P

PEAK

Press

2 Sec

600V CATII

300V CATIII 600A

H

HOLD

OFF

A

A

V

V

ZERO

P H

A

600V CATII

300V CATIII

600V

���

P

PEAK

Press

2 Sec

Wrong

Right

AUTO

P H

A

V

Relative Indicator

AC/DC Indicator

Low Battery

Auto Power Off

Digital Reading

Ohm Indicator

Voltage Indicator

Ampere Indicator

Continuity

Data Hold

Peak Hold