Section 4: Service & Maintenance

04/10

30

51205 Rev. 2

5.2. TROUBLESHOOTING

From the following chart, identify the symptom(s) and cause(s) of each malfunction,
solving each issue with an appropriate solution.

SYMPTOM

CAUSE

SOLUTION

Displays stay at
zero

1. Load Cell connections faulty.

2. Instrument faulty.

3. Faulty/bad Load Cell(s).

1. Cable replacement.

2. Service Instrument.

3. Test and replace the Load Cell(s) as

described in

Section 5.3.3.

Erratic Weights

1. Foreign object around load cells

or under platform.

2. Excessive vibration near platform.

3. Instrument faulty.

4. Platform not level within ¼” (3.0°).

5. Surface not smooth enough

(within 1/8”).

6. Faulty/bad Load Cell(s).

1. Clear the area.

2. Remove the vibration source.

3. Service Instrument.

4. Level the platform surface.

5. Find a smoother surface for the

platform.

6. Test and replace the Load Cell(s) as

described in

Section 5.3.3.

Inaccurate
Weights

1. Instrument out of span.

2. Instrument not properly adjusted

to zero.

3. Faulty/bad Load Cell(s).

1. Check and alter per the Instrument

Service Manual.

2. Zero the instrument according to

normal operation procedures.

3. Test and replace the Load Cell(s) as

described in

Section 5.3.3.