Section 4: Calibration
4.1.
– Calibration Procedure
With the power removed from the instrument, remove the jumper, JP3,
inside the instrument. This will permit access to the calibration menu.
Software Versions 1.6 and higher
Press and hold any one of the following keys
,
,
, or
while turning the scale ON. The display will indicate
,
continue to
hold keys until
is shown on the LCD display. Press
key to
scroll to the
menu. Advance to the Calibration procedure below.
- Calibration Procedure
1. Press
key to enter Calibration procedure.
2. The liveload counts value is displayed on the LCD.
3. Press the
key to display the liveload span counts value.
4. Press
key to clear the liveload span value.
5. Place the calibration weight on the scale platform.
! – IMPORTANT -- !
Remember to re-install JP3 to prevent accidental access to the calibration
upon completing the calibration procedure. Legal for Trade applications
require this jumper to be in place. If the JP3 jumper is not in place, it will not
print
.