b. With the platform empty push ZERO, the display will indicate zero.
c. Place the loaded carrier on the platform.
d. Read the Gross weight at the display.
2. Net Weighing Using Keypad Tare
a. Enter the known Tare weight through the keypad.
b. Press ENTER
c. The instrument will display a negative weight value and the NET Indicator will
be ON.
d. Place filled container on platform. The display will show the Net weight of the
material.
3. Net Weighing Using AUTO TARE
a. Place an empty container on the platform.
b. Press the AUTOTARE key.The empty container weight is transferred to tare
memory.
c. The weight display will now be zero and the NET Indicator light will be ON.
d. Place the material in the container and the instrument display will show the
Net weight.
4. Entering Product ID
a. Press lD.
b. Current ID is displayed.
c. Press GROSS/NET to exit and retain displayed ID, or enter new ID (up to 6
digits) using the keypad. Press ENTER.
5. Printing Weight Information
a. In any mode above, Press PRINT when weight on the platform is stable.
6. For Other Operating Features, Refer to Section 2
50515
10
Issue #2 4/04