Fairbanks H90-5200-A Digital Instrument User Manual
Page 33

W
=0(OFF) L(LOW) H(HIGH)
y
=0(disable) 1(enable)
RECALL SETPOINT <STX>RSM<CR><ETX>
The setpoint mode is output
ENTER SETPOINT# <STX>ESN#<CR><ETX>
Where # =1,2,3,4, enables setpoint number
RECALL SETPOINT# <STX>RSN<CR><ETX>
The currently enabled setpoint is output
ENTER SETPOINT <STX>ES#xxxxxx[L/K]<CD><ETX>
Where # = 1,2,3,4
RECALL SETPOINT <STX>RS#[L/K]<CR><ETX>
IF # =A, the setpoint and preact information for all outputs is given.
IF # =1,2,3,4 the setpoint value is output. L(LB) OR K(KG) can
be used to request output in specific weight units.
ENTER PREACT <STX>EP#xxxxxx[L/K]<CR><ETX>
Where # =1,2,3,4 . The preact is entered in current weight
units. Any decimal point must be entered. L(LB) OR K(KG) can
be used to select specific weight units.
RECALL PREACT <STX>RP#[L/K]<CR><ETX>
IF # = 1,2,3,4, the preact value is given. L(LB) OR K(KG) can
be used to request output in specific weight units.
ENTER UNITS <STX>EUxx<CR><ETX>
xx=LB or KG
I. Computer Command Protocol
1. To program the Computer command protocol, press FUNC 3. The
display will show “CAL IO” Press 0, ENTER.
2. The display will show “ECHO x”, where x is 0 or 1. Select 0/1 to
disable/enable retransmission of each command after CR is received. If
the instrument is configured as Computer - Polled with ID, it will not
retransmit the command unless the ID is valid.
3. The display will show “AC x”, where x is 0 or 1. Select 0/1 to
disable/enable transmission of acknowledgement that command was
or was not successfully executed. If the command was carried out
successfully, a “>” character will be transmitted as acknowledgement.
If the command was not executed for whatever reason, a “?” will be
transmitted as a failure indication.
4. The display will return to the normal weight display.
50515
33
Issue #2 4/04