Appendix i: troubleshooting/error codes – Fairbanks 2800 Series Intrinsically Safe Instrument User Manual
Page 49

01/12
49
50647 Rev. 9
Appendix I: Troubleshooting/Error Codes
FAULT
CAUSE
SOLUTION
-UL--
LOADCELL FAILURE
Scale restricted.
Debris under scale.
Scale behind 0.
Load Cell failure.
Correct the bind or restriction.
Remove debris under the
scale/cells.
Replace items removed from
the scale.
Call Fairbanks service.
--oL--
LOADCELL
FAILURE
Overloaded Scale.
Load Cell failure.
Remove weight from the scale.
Call Fairbanks service.
RECHARGE
Battery voltage low
between 5.4 to 5.32VDC.
Recharge the battery.
LoBAtt
Battery voltage low, between
5.3
to 5.12VDC
Recharge the battery
IMMEDIATELY.
err 069
Programming input error.
Recycle the power to
instrument.
likely lc error
0shift
out of range
Load Cell Failure.
Remove weight from the scale.
Re-zero the scale.
Call Fairbanks service.