Specifications – General Tools and Instruments DLM204 User Manual

Page 9

Advertising
background image

To display the maximum light level recorded during the
current measurement session, press the M-H button. Doing
so will cause the term MAX to appear on the top line of the
LCD, above the highest (maximum) value recorded.

To return to displaying real-time light-level readings, press
the M-H button again.

SPECIFICATIONS

Overall Measurement Range: 0 to 20,000 foot-candles

(ft-cd)/0 to 200,000 lux

Individual Ranges/Resolutions: 0 to 20 ft-cd/0.01 ft-cd;

0 to 200 ft-cd/0.1 ft-cd; 0 to 2,000 ft-cd/1 ft-cd; 0 to
20,000 ft-cd/10 ft-cd; 0 to 200 lux/0.1 lux; 0 to 2,000 lux/
1 lux; 0 to 20,000 lux/10 lux; 0 to 200,000 lux/100 lux

Measurement Accuracy: ±3% of reading for standard

incandescent lamp operating at 2856°K; ±6 to 8% for
other visible light sources (incandescent, fluorescent,
high-pressure sodium and metal-halide lamps)

Angular Deviations from Cosine Characteristics:

±2%@30°, ±6%@60°, ±25%@80°

Ranging Type: Manual only

Display Type & Size: 2 in. (50mm) diagonal LCD with 1 in.

(25mm) high digits

Display Resolution/Refresh Rate:

2000 counts/2.5 times/second

Light Sensor Type: Silicon photodiode

Environmental Classification: Pollution Degree 2

(offices, laboratories and test stations)

Operating Temperature:

14° to 122°F (-10° to 50°C)@<80%RH

9