COMBUSTIBLE GAS DETECTOR

12

ƒ

Return the sensitivity switch to HIGH before

searching for additional leaks.

ƒ

When finished leak testing, turn OFF the instrument

and store it in a clean place. Protect the leak detector
from possible damage.

8. REPLACING NEW SENSOR

The sensor has a limited operative period. Under normal
operation, the sensor should be replaced once a year.
Exposing the sensor to high concentration of coolant
(>30000ppm) will shorten its life cycle rapidly. It is
important to make sure that sensor surface is free from
water droplets, vapor, oil, grease, dust and all other forms
of contaminants. For proper operation, the sensors must
be replaced annually or when contaminated.

WARNING!

When replacing sensor, the

worn-out sensor may be HOT!!

(1) Remove cone cap cover from the tip of the probe.
(2) Pull out old sensor and insert the new sensor into the

plug (see below fig.2).

(3) Seal the cap cover over the plug.