Zero adjustment – General Tools and Instruments PWM301 User Manual

Page 11

Advertising
background image

5. ZERO ADJUSTMENT

Power ON the meter and do not make the measurement,

if the display exist certain value ( not zero ), then it can

according the following procedures to make the Zero

Adjustment :

1)Power Off the meter, press the " Hold Button " ( 3-4,

Fig. 1 ) continuously ( not release ), then press the

" Power Button " ( 3-3, Fig. 1), until the " Display " (

3-2, Fig. 1) show the value, then release the both

buttons.

2)* Press the " Hold Button " ( 3-4, Fig. 1 ) once, the

Display will show the " HOLD " indicator and the value.

* Press the " REC button " ( 3-5, Fig. 1 ) once, the

Display will show the " CAL " text and go to make the

Zero adjustment, final the Display will show End " text

and finish the Zero Adjustment procedures, the

Display will show the zero value.

9