Error codes – Robinair 16910 ID Plus Refrigerant Identifier User Manual

Page 20

Advertising
background image

18

Error Codes

Error Codes and Corrective Actions

The instrument is supplied with self-diagnostic software to guide you through problems

that may occur in the field. The software supples three codes that direct you toward

corrective measures. Should the corrective measures not clear the problem, contact

Technical Services for assistance.

ERR.1

Signifies unstable readings due to inconsistent sample supply or interference from

strong electrical fields.

Corrective Action for Error 1:

1. If the error code occurs during a sampling mode, verify the sample hose is receiving a

sample of at least 9 psig. Also, verify the sample exhaust and air intake ports are clear

and not obstructed in any manner.

2. If this error occurs during a calibration mode, verify the purge vent hose is not directed

towards the instrument, and the air intake and sample exhaust ports are clear and not

obstructed.

3. The unit must be kept away from sources of strong electrical fields (such as large

compressors). Move the instrument 3 to 5 feet (1 to 1.5 meters) away from such

sources and try again.

4. If the instrument is being used in an environment outside the operating temperature

range of 39° F to 122° F (4° C to 50° C), allow the instrument an additional warm-up

period of 15 minutes after power-up before use.

5. To clear the error code, unplug the instrument. The error code will be cleared at the

next power-up.