Error codes – Robinair 16910 ID Plus Refrigerant Identifier User Manual

Page 21

Advertising
background image

19

Error Codes

ERR.2

Signifies internal sensor failure due to temperature extremes.

Corrective Action for Error 2:

1. Allow the unit to stabilize at room temperature for at least 30 minutes.
2. To clear the error code, unplug the instrument. The error code will be cleared at the

next power-up.

ERR.3

Signifies a calibration fault.

Corrective Action for Error 3:

1. Verify the purge vent hose is not directed towards the instrument, and the air intake

port and the sample exhaust ports are not obstructed.

2. If the instrument is being used in small, closed areas, refrigerant vapors can build up

in the surrounding atmosphere. These vapors are drawn into the air intake port during

calibration and will cause a fault. Move the instrument to a location with fresh, ambient

air free of refrigerant vapors. Use the instrument in locations that provide adequate

ventilation to prevent the accumulation of refrigerant vapors.

3. To clear the error code, unplug the instrument. The error code will be cleared at the

next power-up.