Unit description – Robinair 16910 ID Plus Refrigerant Identifier User Manual

Page 7

Advertising
background image

5

Unit Description

The instrument provides the option to purge sample refrigerant sources of air

contamination down to levels set by the user. Air purging is completely automatic

and requires no intervention by the user other than telling the instrument to begin the

purging process.

The instrument is supplied complete with R-12 and R-134a sample hoses, an R-134a

adapter fitting to permit sampling of ACME-ported cylinders, a purge vent hose, a power

cord, and all required plumbing housed within a rugged portable case.

Refrigerant Identification

Sample gas is admitted into the instrument through the supplied sample hose and

is presented to the sensing device. The instrument provides the user with direct

percent-by-weight concentrations of R-12, R-134a, R-22, and hydrocarbons. If the

sample is determined to be pure R-12 or pure R-134a, the instrument also provides a

direct readout of the weight percentage of air within the sample. Note that the instrument

does not consider air to be a contaminate, since it can be removed by refrigerant recycling

equipment or by the purging feature of the instrument itself. Since air is not considered a

contaminate, it is possible to read 100% R-12 plus 5% air. The instrument only considers

the weights of refrigerant contaminates in the total mixture.

The instrument interfaces with the user through the use of a 2-line, 16-character,

alphanumeric display, status indicator lamps, pushbutton communication switches, and

an alarm. Alarm indications are provided to alert the user of instrument fault conditions,

potentially flammable refrigerant presence, and contaminated refrigerant presence. Direct

percent-by-weight concentrations of sample refrigerant are provided on the display, as

well as user directions and prompts. A printer port is supplied with the instrument that

interfaces with most parallel port printers.

Advertising