The User has the ability to change the DSO original factory settings via the “Power Options
Properties” menu in Windows by following the path: Settings – Power Options. Its important to
note that the Windows Power Option named “Standby” provides control of only the DSO’s
computer subsystems (CPU, hard drive, etc.) and does not affect the other subsystems within the
DSO. In general, these other subsystems remain fully powered. For additional information on
setting these Power Options, see the Windows Help menu or other related technical
documentation. In terms of control buttons, this DSO uses only a power button/switch and
therefore references to a sleep button are not applicable.

The DSO can always be placed in the Standby state (LED Off) Power Off (except for some
“housekeeping” circuits) by pressing and holding in the Standby toggle switch for approximately 5
seconds.

Fuse Replacement

WARNING

For continued fire protection at all line
voltages, replace fuse with the specified type
and rating only. Disconnect the power cord
before replacing fuse.

Set the DSO Standby (power) switch to
Standby mode (LED off) and disconnect the
power cord before inspecting or replacing the
fuse. Open the black fuse holder (located at
the rear of the instrument directly to the right of
the AC inlet) using a small, flat-bladed
screwdriver. Remove the old fuse, replace it
with a new 5x20 mm IEC 127 Time Lag (“T”
rated) 10 A / 250 V fuse, and reinstall the fuse
holder.

Calibration

The recommended calibration interval is one year. Calibration should be performed by qualified
personnel only.

Cleaning

WARNING

No operator serviceable parts inside. Do not
remove covers.

Refer servicing to qualified personnel.

Clean only the exterior of the X-Stream DSO,
using a damp, soft cloth. Do not use chemicals
or abrasive elements. Under no circumstances
allow moisture to penetrate the instrument. To
avoid electrical shock, unplug the power cord
from the AC outlet before cleaning.