You can find out the scope's software and hardware configuration as follows:

1. In the tool bar, touch Utilities.

2. In the dialog area, touch Status.

Default Settings

You can reset the scope to default settings by simply pressing the D

EFAULT

S

ETUP

button on the

front panel. The default settings are as follows:

Vertical

Timebase

Trigger

50 mV/div

50.0 ns/div

DC, C1, 0 mV level

0 V offset

10.0 GS/s

edge trigger

positive edge

0 s delay

Auto trigger mode

Adding a New Option

To add a software option you will need a code to enable the option. Call LeCroy Customer
Support at 1-800-553-2769 to place an order and receive the code.

Restoring Software
Restarting the Application

Upon initial power-up, the scope will load the application software automatically. If you exit the
application and want to reload it, touch the scope shortcut icon on the desktop.

If you minimize the application, touch the scope logo button in the task bar at the bottom of the
screen, then touch Maximize.
Restarting the Operating System

If you need to restart the Windows® operating system, you will have to reboot the scope by
pressing and holding the power switch in for 10 seconds, then turning the power back on.