7. Touch

inside

the

Limit field and enter a value, using the pop-up numeric keypad. This

value takes the dimensions of the parameter that you are testing. For example, if you are

testing a time parameter, the unit is seconds. If you chose either WithinDeltaPct

or

WithinDeltaAbs

from the Condition menu, you also have the choice of setting the

limit by means of the statistical buttons at the bottom of the "ParamCompare" dialog:

Comparing Dual Parameters

1. Touch

Analysis in the menu bar, then Pass/Fail Setup... in the drop-down menu.

2. Touch

a

Qx tab; a setup dialog for that position will open.

3. Touch

inside

the

Condition field in the main dialog and select DualParamCompare

.

4. Touch

inside

the

Source1 and Source2 fields and select a source from the pop-up menu.

5. Touch inside the "ParamCompare" mini-dialog field and select a source from the pop-up

menu.

6. Touch

inside

the

Compare Values field and select All or Any from the pop-up menu.

By selecting All, the test will be true only if every waveform falls within the limit that you will
set. By selecting Any, the test will be true if just one waveform falls within the limit.

7. Touch

inside

the

Condition field in the "ParamCompare" mini-dialog and select a math

operator from the pop-up menu: