Saving the calibration data
The calibrations can be saved on the hard disk in the instrument. Generally, as long as the same
cables, connectors, and sampling modules are used, the calibrations can be re-used. Note that the
module calibration ( in the channel menu) is required whenever the instrument is powered up even
if you are using stored TDR calibrations. In addition to the calibrations coefficients, the deskew and
time base settings are also stored.
1. Enter the desired file name in the Calibration Data control.
2. touch the Browse button to select the directory in which to store the calibrations
3. touch the Save button
to recall calibration
1. touch the Calibrate button in any one of the channel menus if the instrument has been turned on
2. touch the Browse button and select the desired file name
3. touch the Recall button