Specifications, Nominal characteristics, Warranted characteristics – Teledyne LeCroy CP150 User Manual
Page 20: Typical characteristics
![background image](/manuals/353486/20/background.png)
CP150 Current Probe
14
922175-00 Rev A
Specifications
Nominal Characteristics
Nominal characteristics describe parameters and attributes which are guaranteed by design, but do not
have associated tolerances.
Maximum Continuous Input Current
150 A
rms
(Refer to Figure 3)
Maximum Peak Current
300 A
peak
, noncontinuous
Insertion Impedance
Refer to Figure 4
Intended Output Load
1 MΩ
Maximum Permitted Circuit Voltage
600 V Cat II, 300 V Cat III (Insulated conductor)
Warranted Characteristics
Warranted characteristics describe parameters which have guaranteed performance. Unless otherwise
noted, tests are provided in the Performance Verification Procedure for all warranted specifications.
Guaranteed at 23 °C ±3 °C (73 °F ±5 °F) after power has been applied for 30 minutes.
Low Frequency Accuracy
±1.0% of reading ±100 mA
Typical Characteristics
Typical characteristics describe parameters which do not have guaranteed performance; however are
representative of the average performance from a sample of several probes. Tests for typical
characteristics are not provided in the Performance Verification procedure.
Sensitivity*
200 mA/div* to 100 A/div.
Output voltage
0.01 V/A
Sensitivity Temperature
±2% or less. (from 0° to +40 °C,
Coefficient
32 °F to 104 °F)
Noise
Equivalent to 2.5 mA
rms
or less
(BW of measuring instrument: 20 MHz)
Bandwidth
DC to 10 MHz
Rise Time
≤ 3.5 ns
* Lower sensitivity may change with different oscilloscope models.