Specifications, Nominal characteristics, Warranted characteristics – Teledyne LeCroy CP150 User Manual

Page 20: Typical characteristics

Advertising
background image

CP150 Current Probe

14

922175-00 Rev A

Specifications

Nominal Characteristics

Nominal characteristics describe parameters and attributes which are guaranteed by design, but do not
have associated tolerances.

Maximum Continuous Input Current

150 A

rms

(Refer to Figure 3)

Maximum Peak Current

300 A

peak

, noncontinuous

Insertion Impedance

Refer to Figure 4

Intended Output Load

1 MΩ

Maximum Permitted Circuit Voltage

600 V Cat II, 300 V Cat III (Insulated conductor)

Warranted Characteristics

Warranted characteristics describe parameters which have guaranteed performance. Unless otherwise
noted, tests are provided in the Performance Verification Procedure for all warranted specifications.

Guaranteed at 23 °C ±3 °C (73 °F ±5 °F) after power has been applied for 30 minutes.

Low Frequency Accuracy

±1.0% of reading ±100 mA

Typical Characteristics

Typical characteristics describe parameters which do not have guaranteed performance; however are
representative of the average performance from a sample of several probes. Tests for typical
characteristics are not provided in the Performance Verification procedure.

Sensitivity*

200 mA/div* to 100 A/div.

Output voltage

0.01 V/A

Sensitivity Temperature

±2% or less. (from 0° to +40 °C,

Coefficient

32 °F to 104 °F)

Noise

Equivalent to 2.5 mA

rms

or less

(BW of measuring instrument: 20 MHz)

Bandwidth

DC to 10 MHz

Rise Time

≤ 3.5 ns

* Lower sensitivity may change with different oscilloscope models.