WaveLink Medium-Bandwidth (8-13 GHz) Differential Probe

Performance Verification

This procedure can be used to verify basic operation and some of the warranted characteristics of
the WaveLink Differential Probe.

The recommended calibration interval for differential probes is one year. Test results can be
recorded on a photocopy of the "Performance Verification Test Record" provided at the end of this
procedure.

Performance Verification can be completed without removing the probe covers or exposing the user
to hazardous voltages. No adjustments are provided.

In the unlikely event a probe fails performance verification, it can be sent back to the local service
center or the factory. For information on returning the probe, refer to "Returning a Product for
Service" on page 36.

If the probe package includes multiple probe interconnect leads, the best practice is to complete the
entire procedure for each probe interconnect lead. If more than one interconnect lead is being
verified, copy and fill out a separate test record for each platform/cable assembly, amplifier module,
and interconnect lead.

If the probe package includes probe interconnect leads that are not serialized to the amplifier
module and platform/cable assembly, then performance may not be verifiable.

NOTE: It is recommended that the Quick Functional Check be performed prior to the Performance
Verification Procedure to assure all other non-warranted functions perform as specified.