Teledyne LeCroy WaveLink Differential Probe Series (13-25 GHz) User Manual
Page 62

WaveLink Series Differential Probe (13-25 GHz)
56
WL-HBW-A-OM-E Rev B
8.
At this point, the Calibrator must be set up before performing the functional tests. Select
Utilities → Utilities Setup...
from the menu bar.
9.
Touch the
Aux Output
tab as follows:
10.
Touch the
Square
button to obtain a square wave output signal.
Auxiliary output setup
NOTE: TTL is processed internally (and the Aux Output dialog shows a TTL checkbox) for the
following oscilloscope models: WaveMaster 5000, 6000, SDA 4000, 5000, and 6000.
11.
Set the amplitude to 1 Volt, Frequency to 1.00 kHz.
12.
Verify the screen indicates the proper settings in the Aux Output fields.
13.
Set the probe sensitivity to 200 mV/div.
Connect the + tip of the interconnect lead module to the center connector of the calibrator
output signal, and the – lead to the shell (ground) of the connector. If necessary, readjust the
tip spacing.