Verify the screen shows a square wave (you may need to adjust the channel offset). Refer to
the following screen-shot. If no square wave is shown, the
+
channel of the probe may be
faulty.
Square wave output signal
16.
Make desired adjustments to your oscilloscope settings and obtain a stable display.
17.
Adjust the tips so both touch or are almost touching.
18.
Connect both tips to the calibrator output signal.
19.
Verify a straight line is shown centered on the screen. There should be no vertical deflection,
to indicate good CMRR.
20.
One of the channels may be at fault if a square wave or part of a square wave is shown.
This concludes the functional tests of the WaveLink High Bandwidth Differential Probe.