Performance verification tests, Instruction manual, Cw gain – Teledyne LeCroy OE695G User Manual
Page 15

Instruction Manual
922562-00 Rev A
9
Performance Verification Tests
CW Gain
Required Equipment
OE695G and compatible oscilloscope
CW laser source at the desired optical wavelength
Variable optical attenuator (or use variable output level of CW source)
Optical power meter calibrated for the optical wavelength of the source
Multimode optical fiber patch cable
Procedure
1. Connect the OE695G to a compatible oscilloscope channel and allow to
fully warm-up.
2. Configure the following settings:
a. Recall Default setup
b. Enable the channel with the OE695G, disable others
c. Configure the active channel for the optical wavelength
matching the laser source
d. Create a measurement for the mean value of the channel
e. Enable measurement statistics.
3. Connect the optical attenuator (if used) input fiber to the laser source or
connect a fiber patch cable to the output of the laser source. Before
mating, carefully clean the fiber connections with an appropriate fiber
cleaning cloth.
4. Set the attenuator/laser for an output level of approximately
60 uW.
5. Connect the patch/attenuator output cable to the optical power meter
and note the Input Power Measurement: _________ uW
6. With the OE695G optical input cover closed, perform a dark calibration
(page 6). Press Clear Sweeps on the oscilloscope front panel. From the