Performance verification tests, Instruction manual, Cw gain – Teledyne LeCroy OE695G User Manual

Page 15

Advertising
background image

Instruction Manual

922562-00 Rev A

9

Performance Verification Tests

CW Gain

Required Equipment

OE695G and compatible oscilloscope

CW laser source at the desired optical wavelength

Variable optical attenuator (or use variable output level of CW source)

Optical power meter calibrated for the optical wavelength of the source

Multimode optical fiber patch cable

Procedure

1. Connect the OE695G to a compatible oscilloscope channel and allow to

fully warm-up.

2. Configure the following settings:

a. Recall Default setup

b. Enable the channel with the OE695G, disable others

c. Configure the active channel for the optical wavelength

matching the laser source

d. Create a measurement for the mean value of the channel

e. Enable measurement statistics.

3. Connect the optical attenuator (if used) input fiber to the laser source or

connect a fiber patch cable to the output of the laser source. Before
mating, carefully clean the fiber connections with an appropriate fiber
cleaning cloth.

4. Set the attenuator/laser for an output level of approximately

60 uW.

5. Connect the patch/attenuator output cable to the optical power meter

and note the Input Power Measurement: _________ uW

6. With the OE695G optical input cover closed, perform a dark calibration

(page 6). Press Clear Sweeps on the oscilloscope front panel. From the

Advertising