Current measurement calibration board – Teledyne LeCroy Conquest User Manual - Users Manual User Manual
Page 182

Current Measurement (Optional)
172
Current Measurement Calibration Board
The Current measurement calibration board is provided to verify the current
measurement functionality of Conquest M2.
Figure 133 Current Measurement Calibration Board
Table 6 shows the connections to the Analyzer, calibration board switch
settings and pass/fail criteria for the tests.
Table 6 Measurement Calibration Parameters
* This test must be run twice to change from 5A to 10A range.
** The calibration board is not used for this test. Only connect the host
exerciser port to the Main Analyzer Port.
*** There is a ±3 µC tolerance on the measurements limited by the
components on the calibration board.
Current Test
Pass
Limit
Connection
S1-1 S1-2 S1-3 S1-4
Conquest
M2
Measure-
ment
Inrush 1 (5A range)
50µC
Inrush
ON
OFF OFF OFF
40µC***
Inrush 2* (10A
Range)
50µC
Inrush
ON
ON
OFF OFF
50µC***
Unconfigured
100mA
Analyzer/Exerciser OFF OFF
ON
OFF
66mA
Suspend
500µA
Analyzer/Exerciser OFF OFF OFF
ON
107µA
Operating
500mA
Analyzer/Host
OFF OFF
ON
OFF
66mA
VBus**
5.25V
Analyzer/Exerciser
N/A
N/A
N/A
N/A
5.176V