Protocol Analysis
62
Counters
The sequencer incorporates two counters each of which you can configure as
a timer or an event counter, with the following commands:
•
Set_Timer1
•
Set_Counter1
•
Enable_Timer1
•
Enable_Counter1
•
Set_Timer2
•
Set_Counter2
•
Enable_Timer2
•
Enable_Counter2
Set the counters with the Set_Timer and Set_Counter commands, with
corresponding timer or counter values in any state.
Whenever a state transition occurs, the timer or counter stops, unless an
Enable_Timer or Enable_Counter command is set in the next state.
Start Exerciser
You can start the exerciser at “State0” only. To start the exerciser, click the
ellipses next to “State0”. Note, however, that you must program the exerciser
prior to running the project to get meaningful results. For instructions on
programming the exerciser, see page 67.
External Outputs
You can set the eight external output bits on the output connector at any state
by clicking the ellipses next to the sate, choosing External_Bits from the
drop-down list for that state, and then entering a bit pattern. See page 15 for
external output connector pin assignment.