ATO-400 / ATO-600 Operating Procedures

37

Table 12.0 Calibration Test Procedure (Continued)

STEP

ACTION

ATO DISPLAY

12-9

Connect the 100 micro-ohm shunt to ATO.
Press “ENTER” key to start test.

12-10

The ATO will measure the 100

?

-ohm shunt

using 10A test current.

12-11

Observe test results. Press the “ENTER”
key to continue.

12-12

The ATO will measure the 100

?

-ohm shunt

using 100A test current.

12-13

Observe test results. Press “ENTER” key to
continue.

12-14

Connect the 500 micro-ohm shunt to ATO.
Press “ENTER” key to start test.

12-15

The ATO will measure the 500

?

-ohm shunt

using 10A test current.

12-16

Observe test results. Press “ENTER” key to
continue.

CONNECT 100

? -OHM

0.1% SHUNT NOW… .
“ENTER” TO CONTINUE

CURRENT: 10.0 AMP
BURN IN: 02 SEC
100

? -OHM TEST

FINAL RESISTANCE
I= 10.0 AMPS
100.1 MICRO-OHM
ANY KEY TO CONTINUE

CURRENT: 100.0 AMP
BURN IN: 02 SEC
100

? -OHM TEST

FINAL RESISTANCE
I= 100.0 AMPS
100.1 MICRO-OHM
ANY KEY TO CONTINUE

CONNECT 500

? -OHM

0.1% SHUNT NOW… .
“ENTER” TO CONTINUE

CURRENT: 10.0 AMP
BURN IN: 02 SEC
500

? -OHM TEST

FINAL RESISTANCE
I= 10.0 AMPS
500.1 MICRO-OHM
ANY KEY TO CONTINUE