CT-6500 Series 2 Operating Instructions

57

9.14 Transducer Self-Test

The operator can check the transducer electronics by connecting the transducer to the CT-

6500 (D1 or D2 or D3) and running a “Test Transducer” diagnostics.

Table 19.0 Performing a Transducer Self Test

STEP OPERATION

ACTION

DISPLAY

1 SELECT

DIAGNOSTIC.

PRESS # 4 KEY.

1. SLOW-CLOSE TEST

2. CHECK HOOKUP

3. TEST TRANSDUCER

4. NEXT PAGE

2

SELECT TRANSDUCER.

PRESS # 3 KEY.

TRANSDUCER TEST

1= 0.00 IN / 000.0

°

2= 0.00 IN / 000.0

°

3= 0.00 IN / 000.0

°

3

MOVE TRANSDUCER SLIDER TO A

KNOWN DISTANCE & OBSERVE

DISPLAY

MOVE

TRANSDUCER

SLIDER

TRANSDUCER TEST

1= 5.00 IN / 360.0

°

2= 0.00 IN / 000.0

°

3= 0.00 IN / 000.0

°

2

RETURN O MAIN MENU

PRESS # STOP

KEY.

1.RUN TEST 06/10/07

2.GET RESULTS 20:10:01

3.SETUP

4.DIAGNOSTIC

NOTES

1. After selecting the transducer test mode, move the transducer slider and observe the
display on the CT-6500 LCD.

2. If the slider is moved 1.00 inch, the CT-6500 LCD should display 1.00 inch.

3. Press the “STOP” key to return back to the main menu.