For further information about the functioning of

the clinical AED Plus, see the AED Plus Operator’s Guide

and Administrator’s Guide.

The Remote Controller lights indicate the current training

settings while the Trainer2 simulates the actual AED Plus behavior.

The Remote Controller will turn itself off after 10 seconds.

The Remote Controller will remember the last scenario setting which

reappears the next time the View/Enter Button is pressed.

NOTE

SETUP and TEST

1

SELECT SCENARIO SETTINGS

2

Scenario 3: SHOCK - NO SHOCK

Scenario 2: NO SHOCK - SHOCK

Scenario 1: SHOCK - SHOCK

Manual Scenario Mode

CONTROLS / INDICATORS

Handheld Remote Controller:

PUSH HARDER prompt

(only when CPR Metronome is beeping).

NO SHOCK advised on next heart analysis

(manual mode only).

Pads attached/not attached to victim’s bare chest

(toggle-switch).

View/Enter scenario

Show current scenario selection (press once)

Start scenario (press again)

Move Up to next scenario.

Scenario 4: NO SHOCK - NO SHOCK

Set-up Controls:

Training Controls:

GOOD COMPRESSIONS prompt

(only when CPR Metronome is beeping).

Pause (resume) current scenario.

End current scenario.

NOTE

Install 4 “D” alkaline batteries in the AED Plus Trainer2.

Trainer2 unit:

Action:

Press On/Off Button to turn Trainer2 on.

Green light turns on, indicating Trainer2 is

ready for use.

Remote Controller:

Install 2 “AA” batteries in handheld remote.

Turn the Remote Controller on by pushing the

View/Enter Button.

Point Remote Controller at unit and press

View/Enter Button again. Green light on Remote

Controller will show Manual Scenario Mode is

selected.

Manual Scenario Mode
View/Enter Button

Trainer2 unit:

Action:

Turn Trainer2 unit off (hold down On/Off Button).

Setup test is now complete. Trainer2 unit and

Remote Controller are ready for use.

Trainer2 unit:

Action:

Press the On/Off Button to turn on the Trainer2.

Action:

Remote Controller:

Action:

Use Move Up Button to select desired scenario.

Use CPR Metronome On/Off Button to turn

Real CPR Help

®

(beeping) on or off for this

scenario.

Press the View/Enter Button once to transmit the

selected Training Scenario to the Trainer2 unit.

CPR PROMPTS

Remote Controller:

Action:

The Instructor uses the Remote Controller to control prompts during a

Training Scenario’s CPR period. The instructor can use the following buttons

to simulate the use of an AED Plus defibrillator with Real CPR Help.

Remote Controller:

Action:

In all scenarios (manual and automatic), the rescue

can be paused by pressing the Pause/Resume

Button once. Press it a second time to resume the

scenario.

Trainer2 unit:

Action:

The Push Harder and Good Compressions

Buttons are only valid during the CPR intervall.

The red light on the Trainer2’s status window emits

one long flash if the prompt is currently valid.

If the Push Harder or Good Compressions

Buttons are pushed when the Trainer2 is not

beeping, the unit’s red light emits three quick

flashes, indicating that the prompt is currently

invalid.

In all scenarios (manual and automatic), the rescue

can be ended by pressing the End Button once.

NOTE

On
Off

On
Off

On
Off

The buttons above the line on the handheld Remote Controller are Set-up

Controls, which the instructor uses to select a training scenario (an LED lights

next to the selected scenario). The buttons below the line are Training Controls,

which the instructor uses to introduce events into the training scenario.



Good Compressions Button

Accurately imitating the student’s AED Plus unit requires

instructors be familiar with the protocol currently followed by the
AED Plus for PUSH HARDER and GOOD COMPRESSIONS prompts.

NOTE

Pressing this button during the CPR period causes

the Trainer2 to issue the GOOD COMPRESSIONS voice

prompt. Intructors should press this button to provide

students with positive feedback when they respond

to the PUSH HARDER prompt by improving the depth

of chest compressions


Push Harder Button

Pressing this button during the CPR period causes

the Trainer2 to issue the PUSH HARDER voice prompt.

Instructors should press this button when a student's

chest compressions are too shallow.

SCENARIO CONTROLS

3

Remote Controller:

Action:

VF (Ventricular Fibrillation-shock advised) is the

default simulated heart rhythm in Manual Scenario.

Analysis - VF

For the next analysis to result in NO SHOCK ADVISED,

push the No Shock button.

NOTE

Once the No Shock button has been pushed, the next

simulated heart analysis will advise NO Shock. This setting cannot be

reversed for the next analysis.

In all scenarios the Pads Button must be pressed

once to proceed with analysis. If pressed again, it

will begin to prompt for attaching pads.

Press to turn on Trainer2.

Press again to start current scenario

(as selected using Remote Controller).

Press and hold down to turn off Trainer2.

ZOLL AED Plus

®

Trainer2:

Green light remains on = Trainer2 is on and ready for use.

Green light off = Trainer2 is turned off or has no battery power.

Red Light steady = Last command from Remote Controller accepted.

Red Light 3 flashes = Last command from Remote Controller is invalid.

On
Off

CPR Metronome On/Off Button

Enables the CPR metronome to beep for the selected

Training Scenario. The LED above the button lights to

indicate that the CPR metronome is enabled. You should

enable the CPR metronome whenever CPR-D padz

®

are connected to the Trainer2.

When a CPR period occurs in the Training Scenario, the

CPR metronome begins to beep at 60 compressions

per minute (CPM), then automatically accelerates to

100 CPM after a few seconds. Metronome beeps are

independent of the student's actual compression rate

and continue until either the CPR interval ends or the

the instructor ends the scenario.

Press the View/Enter Button once to turn on the

Remote Controller.

CPR Prompts