ZOLL AutoPulse Rev D Quick Reference Guide User Manual

For adult, Non-traumatic cardiac arrest, Troubleshooting

Advertising
background image

1. Power up AutoPulse

2. Close chest bands

3. Press CONTINUE (green button)
4. Press START (green button) to

begin compressions

To pause or stop operation

press STOP (orange button)

• Remove ALL clothing from torso

(both front and back) to ensure

skin-to-platform contact

• Align armpits onto yellow line

on Platform

Do not twist the

chest bands

Maintain chest bands at

90 degrees to Platform

and free of obstructions

For Fault/User Advisory

• Lift up and fully extend both chest bands.
• Check both lateral and vertical patient alignment.
• Verify that chest bands are not twisted,

are 90 degrees to the Platform and are

free of obstructions.

• Press RESTART (green button) and follow

on-screen instructions to begin compressions.

If you cannot rectify problem immediately open chest

bands and revert to manual

CPR

.

Troubleshooting

Follow all local protocols
and procedures.
Review User Guides and
complete in-service training.

LifeBand

®

instructions on other side

P/N 10596-001 Rev. 10

AutoPulse

®

Quick Reference Guide

For Adult

(>–18 years)

Non-Traumatic Cardiac Arrest

Maximum Patient Weight 300 lbs.