Lifeband, Load-distributing band, Removing lifeband installing new lifeband – ZOLL AutoPulse Rev D Quick Reference Guide User Manual

Page 2

Advertising
background image

LifeBand Cover Plate

Tail End

Head End

Band Clip

Guide Plate

2

1

2

1

AutoPulse is a trademark of ZOLL Circulation
ZOLL Circulation ©2009 Printed in U.S.A.
P/N 10596-001 Rev. 10

Removing LifeBand

Installing new LifeBand

• Place AutoPulse face down.
• Lift hinged skirts, pinch 4

locking tabs and remove

Cover Plate.

• Grasp the Band with the

thumb and index finger of

both hands. Push in the

Guide Plate, using both

middle fingers and, pull up

the Band to remove the clip

from the shaft.

• Match arrow on the Cover

Plate with arrow on Platform.

• Insert Head end of Band

Clip into slot. Press Tail

end of Band Clip into Guide

Plate slot and feel for "click."

Rotate shaft in either direction

to verify Band Clip is seated

in slot. Ensure that Band is

not twisted.

• Snap Cover Plate in place and

flip down hinged skirts.

• Important: power on AutoPulse. If

a Fault/User Advisory is displayed,

check installation of the Band Clip

into the Drive Shaft slot.

LifeBand

®

Load-distributing Band

2

1

2

1

AutoPulse deployment on other side

AutoPulse

®

Quick Reference Guide

0344

Inquiries to:

ZOLL Circulation
650 Almanor Avenue
Sunnyvale, CA 94085 U.S.A.

t: +1.800.321.4CPR (4277)
t: +1.408.541.2140
f: +1.408.541.1030

EU Authorized Representative
ZOLL International Holding B.V.
Newtonweg 18
6662 PV ELST
The Netherlands

t: +31 481 366410

Advertising