4 magnetic contact (option /gm1 to /gm5), 4 magnetic contact (option /gm1 to /gm5) -4, Connection to transmitter relay – Yokogawa RAGN User Manual

Page 18

Advertising
background image

<4. INSTALLATION

4-4

IM 01R01B10-00E-E 2nd edition: November 01, 2011-00

All Rights Reserved. Copyright © 2011, Rota Yokogawa

4.4 Magnetic contact (Option /GM1 to /GM5)

a) Loose the nut at the guide sleeve.
b) If 2 limit switches were ordered install the Max-contact in the top position and the Min- contact

in the low position (see print on housing).
c) Put the limit switch from the outer side on the guide rail of the Rotameter.
d) Adjust the distance between limit switch to tube to 1 mm; check function and correct if necessary.
e) Fix limit switch with the nut to guide rail.

Optional connection box /GD1

Limit switch /GM1 or /GM2

Float with magnet

Nut

F45.EPS

GRENZWERTSCHALTER GM

LIMIT SWITCH GM

CONTACT LIMITE GM

ROTAMETER RAGN with option /GMx (x = 1 ... 5)

EN 60947-5-6 (NAMUR)

MAINS / NETZ / TENSION

OPTION: /W4B 24VDC

KFD2-SR2-Ex.W

OPTION: /W2B 230VAC

KFA6-SR2-Ex.W

9

8

1

1

1

7

1

1

1

3

4

6

+

-

+

-

LIMIT/ GRENZWERT/ LIMITE

MAX

MIN

Ä

TRENNSCHALTVERST RKER

TRANSFORMER ISOLATED BARRIER

AMPLIFICATEUR SEPARATEUR

(L+

(L-)

EINKANALIGE VERSTÄRKER .......-SR2-EX1.W ANSCHLUSS WIE GRENZWERT "MAX"

ONE CHANNEL TRANSMITTER .....-SR2-EX1.W CONNECTION LIKE LIMIT "MAX"

UN CANAL AMPLIFICATEUR ........-SR2-EX1.W RACCORD COMME LIMITE "MAX"

LIMIT/ GRENZWERT/ LIMITE

Connection to transmitter relay :