<6. EXPLOSION PROTECTED INSTRUMENTS>

6-3

IM 01R01B10-00E-E 2nd edition: November 01, 2011-00

All Rights Reserved. Copyright © 2011, Rota Yokogawa

Marking:

6.2 Magnetic contact (Option /GM1 to /GM5)

• Only trained persons may use the instrument in the industrial area.
• It is forbidden to users to carry out specification changes and other changes at the device. Repairs at

the device aren’t permitted.

• To ensure explosion protection of the GM, a suitable explosion-proof transformer isolated barrier must

be used.

• The maximum ambient temperature and the maximum temperature of the medium, which flows through

the tube, may not exceed 70°C.

• The connection cable may not be exposed to mechanical loads. The maximum axial strength is 30 N.

• Static charge of the GM case has to be avoided. A corresponding warning note has to be attached at

the device.

The limit switch GM is classified according EN 60079-11 chapter 5.7, IEC 60079-11 chapter 5.7 and ANSI/
ISA 60079-11 chapter 5.7 as "Simple Apparatus".
For use in hazardous area the limit switch must be connected to a suitable explosion-proof transformer isolated
barrier, which does not exceed the following maximum values :
Maximum voltage Uo

: 15 V

Maximum current Io

: 50 mA

Maximum power Po

: 187 mW

Classification :
- Type

: GM1 or GM2

- Protection

: intrinsic safe

- ATEX / IECEx

: IIC 2G

- FM

: I, 1, A, B, C, D

- Temperature class

: T6

- Entity parameters

: Ui = 15 V ; Ii = 50 mA ; Pi = 187 mW ; Li ≈ 0 mH ; Ci ≈ 0 nF

WARNING

Rota Yokogaw

Rheinstr. 8

D-79664 Wehr

Type : RI20-10-K

U = 4.5 ... 15V

SN : ..............

Rota Yokogawa
Rheinstr. 8
D-79664 Wehr
Type : RI20-10-K
U = 4.5 ... 15V
SN : ..............

Ex ia IIC T6
PTB 03 ATEX 2111

II 2 G

see certificate
for data