<3. Commissioning>

13

IM 12A01A02-12E

4th Edition : Sep. 3, 2013-00

3. Commissioning

QUICK SET-UP NOTES:

a. pH measurement module
Under “measurement “a selection is made for pH, ORP or pH&ORP. The selected

measurement must be in accordance with the sensor wiring. When rH measurement is

requested pH&ORP must be chosen on this level. The rH must then be selected in the

commissioning menu.

b. SC measurement module
Under “measurement” a selection is made between Conductivity, Resistivity, Concentration

or Concentration plus Conductivity. On this level only Conductivity or Resistivity can be

selected. Settings for Concentration measurement must be done in the commissioning

menu.

c. DO measurement module
Under “sensor type” a selection is made for Galvanic or Polarographic. The selected sensor

type must be in accordance with the sensor wiring. Otherwise the analyzer or sensor can be

damaged.