<1. Wiring and Installation>

4

IM 12A01A02-12E

4th Edition : Sep. 3, 2013-00

1.3.2

ORP Measurement

The ORP measurement uses the same sensor input module as the pH measurement. It is not

uncommon to measure ORP as process variable and a pH Glass electrode as reference. This is

the case with rH measurement and with pH compensated ORP measurement.
Standard reference electrodes and an ORP/REDOX electrode are low impedance sensor inputs

Conventional ORP sensors are connected to the module as follows:

11 Temperature
12 Temperature
13 Reference
14 Solution ground
15 Metal (measure)
16 Shield
17 Shield

FLXA21

REF

TC

Liquid

Earth

ORP

For conventional ORP sensors, the jumpers are placed as illustrated:

16

15

19

17

13

18

14

12

11

PH

Input 1

Input 2

Metal sensor on Input 1.

Normal reference on Input 2.

pH Compensated ORP sensors are connected to the module as follows:

11 Temperature
12 Temperature
13 Glass
14 Solution ground
15 Metal
16 Shield
17 Shield

FLXA21

PH

TC

ORP

Liquid

Earth

For pH Compensated ORP sensors, the jumper is placed as illustrated:

16

15

19

17

13

18

14

12

11

PH

Input 1

Input 2

Metal sensor on Input 1.

pH glass (as reference) on Input 2.