<1. Wiring and Installation>

2

IM 12A01A02-12E

4th Edition : Sep. 3, 2013-00

1.2

Wiring the power supply

FLXA21 is a 2-wire transmitter and can be powered by a DC power supply of max. 40 VDC. The

Power Supply voltage depends on:

• The load resistance: impedance of electronic equipment: typically 250 Ohm
• Number of input modules: 1-sensor measurement or 2-sensor measurement.

One (1) Sensor module (1 input):

16 to 40 V DC (for pH/ORP, SC and DO)

17 to 40 V DC (for ISC)

21 to 40 V DC (SENCOM pH/ORP)

Two (2) Sensor modules (2 inputs):

22.8 to 40 V DC (for pH/ORP, SC and DO)

250

600

1000

0

24.7

18

V - 11.5

0.022

R =

1295

1617

22.8

40

40

Voltage (V)

2-sensor measurement

Load resistance (Ω)

Digital Communication

Range (HART)

250

304

600

516

1000

0

24.7

18

V - 11.5

0.022

R =

1295

17

40

22.86

40

Except SENCOM

Voltage (V)

Load resistance (Ω)

Digital Communication

Range (HART)

18.2 21

Figure 1 Supply Voltage and Load Resistance for

pH/ORP (analog sensor), SC and DO

Figure 2 Supply Voltage and Load Resistance for

ISC and pH/ORP SENCOM sensor

The power supply is connected to the terminals marked with + and – which corresponds with the
polarity of the DC power supply. The shield/ ground is connected to the terminal marked ,
then replace ground wiring cover.