Setup Example 2: Flow Rate Measurement Channel and Alarm

DX2000

Flowmeter

Channel 2

4 - 20 mA

Convert to 1-5 V with a shunt resistor

Setup Item

Description

Number in the Figure

Channel

Use channel 2.

1

Tag

FI-002

2

Input signal

1-5V

3

Input range

0.0 to 500.0 L/H

4

Alarm condition

Output an alarm if the measured value is
less than or equal to 120.0 L/H.
Output destination: Relay contact (I03)

5

(1) Input Range and Alarm

Press

MENU (to switch to setting mode).

Select the

Menu tab > Meas Channel > Range, Alarm.

1

5

4
3

(2) Tag

Select the

Menu tab > Meas Channel > Tag, Memory sample, Alarm delay.

2

Operation complete.