Channel (setting the burnout and rjc), Channel (setting the burnout and rjc) -21, Index – Yokogawa Data Acquisition with PID Control CX2000 User Manual

Page 33

Advertising
background image

2-21

IM04L41B01-65EN

1

2

3

4

Index

Viewing and Creating Setup Data

Memory Timeup

Specify the date and time to save the data in the internal memory to the external storage
medium.

• Timeup type

Time interval to save the data. Set [OFF] to disable the function.

Hour: 1 hour, Day: 1 day, Week: 1 week, Month: 1 month

• Date

Date (1 to 28) or the day of the week (SUN, MON, TUE, WED, THU, FRI, and SAT)

• Time (Hour)

Hours (00 to 23)

Channel (Setting the Burnout and RJC)

Click here (also selectable from [Setting] - [System Mode Setting]).

Set to the positive side (100%).

Copy and paste (see page 4-15).

Tools (see page 4-6)

Resets the range selection.

Channel number

Select the range by dragging & dropping the mouse.

Set the reference junction

compensation to internal or external.

Set to the negative side (0%).

Burnout

For thermocouple (TC) inputs, select how the measurement results are to be handled
when the thermocouple burns out.

RJC Volt (uV)

Internal: Use the DXP’s reference junction compensation function
External: Use an external reference junction compensation function.
Volt(uV): When the type is set to [External], set the compensation value in the range from
–20000 to 20000.

Copying and Pasting Setup Data

The items checked in [Copy Details] can be copied and pasted. For the operating
procedure, see page 2-5.

2.6 Configuring the System Mode

Advertising
This manual is related to the following products: