4 setting commands (setting), 4 setting commands (setting) -10, Sr sets an input range – Yokogawa Button Operated MV2000 User Manual

Page 84

Advertising
background image

4-10

4.4

Setting Commands (Setting)

IM MV1000-17E

4.4 Setting Commands

(Setting)

SR

Sets an input range

To set a channel to skip

Syntax

SR p1,p2<terminator>
p1 Measurement channel number
p2 Setting type (SKIP)

Query

SR[ p1]?

Example Skip channel 001.

SR001,SKIP

Description • You cannot execute this command while the

MV is measuring or computing.

• A channel set to SKIP is not measured.

• Set parameter p1 according to the table in

section 4.3.

To set a channel to voltage, TC, RTD, or ON/OFF

input

Syntax

SR p1,p2,p3,p4,p5<terminator>
p1 Measurement channel number
p2 Input type

VOLT

DC voltage

TC

Thermocouple

RTD

RTD

DI

ON/OFF input

p3 Measurement range
p4 Span lower limit
p5 Span upper limit

Query

SR[ p1]?

Example Set the channel 001 input type to TC type R, the

span lower limit to 0°C, and the span upper limit
to 1760.0°C.
SR001,TC,R,0,17600

Description • You cannot execute this command while the

MV is measuring, computing, or generating a
report.

• Set parameters p1 and p3 according to the

table in section 4.3.

• For parameters p4 and p5, enter a value using

5 digits or less excluding the decimal point.

To set a channel to difference computation

Syntax

SR p1,p2,p3,p4,p5,p6,p7<terminator>
p1 Measurement channel number
p2 Setting type (DELTA)
p3 Input type

VOLT

DC voltage

TC

Thermocouple

RTD

RTD

DI

ON/OFF input

p4 Measurement range
p5 Span lower limit
p6 Span upper limit
p7 Reference channel number (measurement

channel number)

Query

SR[ p1]?

Example Set the channel 010 setting type to differential

computation between channels with the reference
channel set to 001, and set the input type to TC.
Set the measurement range to R. Set the span
lower limit to 10.0°C and span upper limit to 100.0°C.

SR010,DELTA,TC,R,100,1000,001

Description • You cannot execute this command while the

MV is measuring, computing, or generating a
report.

• Set parameters p1 and p4 according to the

table in section 4.3.

• For parameters p5 and p6, enter a value

using 5 digits or less excluding the decimal
point.

To set a channel to scaling

Syntax

SR p1,p2,p3,p4,p5,p6,p7,p8,p9,p10

<terminator>
p1 Measurement channel number
p2 Setting type (SCALE)
p3 Input type

VOLT

DC voltage

TC

Thermocouple

RTD

RTD

DI

ON/OFF input

p4 Measurement range
p5 Span lower limit
p6 Span upper limit
p7 Scaling lower limit (–30000 to 30000)
p8 Scaling upper limit (–30000 to 30000)
p9 Scaling decimal place (0 to 4)
p10 Unit (up to six alphanumeric characters)

Query

SR[ p1]?

Example Convert the DC voltage measured on channel

002 to DC current. Set the input range to 6 V,
the span lower limit to 1 V, the span upper limit
to 5 V, the scaling lower limit to 1.00 A, and the
scaling upper limit to 5.00 A.

SR002,SCALE,VOLT,6V,1000,5000,100,5
00,2,A

Description • You cannot execute this command while the

MV is measuring, computing, or generating a
report.

• Set parameters p1 and p4 according to the

table in section 4.3.

• For parameters p5 and p6, enter a value

using 5 digits or less excluding the decimal
point.

• Set all parameters p7, p8, and p9 or omit all

three parameters.

To set a channel to square root computation

Syntax

SR p1,p2,p3,p4,p5,p6,p7,p8,p9,p10,p11

<terminator>
p1 Measurement channel number

p2 Setting type (SQRT)

p3 Measurement range

p4 Span lower limit

p5 Span upper limit

p6 Scaling lower limit (–30000 to 30000)

p7 Scaling upper limit (–30000 to 30000)

Advertising
This manual is related to the following products: