6. CAN Bus Module (MX118-CAN-M30) Specifications

6.1.Model Name

Model Name

Description

Max. Number of

Channels

Min. Sampling

Interval

Slot Width

MX118-CAN-M30

CAN Bus Data Acquisition

Module

30 ch

10 ms

1–3



6.2.Module Specifications

Number of Channels: 30 ch/module
No. of messages (ID): 30
Sampling interval: 10 ms–60 s
Max. no. of installed modules: Differs by slot width setting (no. of ch used per module)

Slot Width Setting

Max. No. of Ch Used

Max. no. of Modules (per System)

1 10

ch

6

2 20

ch

5

3 30

ch

4


Multi-interval: Allowed. However, if the slot width is 2 or 3, a separate measurement interval cannot be set

for the real and virtual instances of the same module.


Mode switching: Switch between Setting and Measurement modes with hardware dip switch.

Setting mode: CAN bus settings can be entered externally from a PC using a SERIAL (connector).
Measurement mode: CAN bus data can be captured and transferred to the MW100 using a CAN
(connector).

Data range: -30000–30000 (range of data that can be transferred to the MW100)

Set using the upper/lower limit of span.


Scaling computation: The following scaling computation is carried out inside the module.

SpanMin

SpanMin

SpanMax

ValueMin

ValueMax

ValueMin

InputValue

+

)

(


Input Value

: Input CAN bus data (max. 64 bit data)

Value Max

: Upper limit of CAN bus data set by user

Value Min

: Lower limit of CAN bus data set by user

Span Max

: Upper limit of span set by user

Span Min

: Lower limit of span set by user