Configuring the Recorderr

1

2

3

4

Index

2.6 Setting the Items in Setting Mode and the Data

Display Method

Click the

Setting tab. You can also select the item by choosing SET [Regular] Setting

from the

Setting menu.

Chart Speed/Trend Interval

Click here (or choose [Setting] > [SET [Regular] Setting] > [Chart Speed]/[Trend Interval])

Chart speed

Secondary chart speed

Trend interval (dot model)

Chart Speed

• Pen Model

The chart speed can be selected from 82 settings shown below.

Chart speed on the pen model (unit: mm/h)

5

6

8

9

10

12

15

16

18

20

24

25

30

32

36

40

45

48

50

54

60

64

72

75

80

90

96

100

120

125

135

150

160

180

200

225

240

250

270

300

320

360

375

400

450

480

500

540

600

675

720

750

800

900

960

1000

1080

1200

1350

1440

1500

1600

1800

2000

2160

2250

2400

2700

2880

3000

3600

4000

4320

4500

4800

5400

6000

7200

8000

9000

10800 12000

• Dot Model

The chart speed can be set in the range of 1 to 1500 mm/h in 1 mm steps.

Trend Interval (Dot Model)

Auto: The trend recording interval is set according to the chart speed in the range of

10 s to 90 s so that the dots do not overlap.

Fix:

The trend recording interval is fixed to 10 s/6 dots. On models with the
computation function (/M1 option), the dot printing interval varies depending on
the number of measurement and computation channels to be trend recorded.