IN 6/78-L Operating Manual

Data format UPP (Universal Pyrometer Protocol) · 29

7 Data format UPP (Universal Pyrometer

Protocol)

Via interface and suitable communication software or via Test function of the InfraWin
software (see Section 5.4 Basic settings) commands can be exchanged directly with the

pyrometer.
The data exchange occurs in ASCII format with the following transmission parameters:
The data format is: 8 data bits, 1 stop bit, even parity (8,1,e).
The device responds to the entry of a command with: output (e.g. the measuring value) + CR
(Carriage Return, ASCII 13), to pure entry commands with ok + CR.
Every command starts with the 2-digit device address AA (e.g. "00"). This is followed by 2 small
command letters (e.g. "em" for level of emissivity

e

), finished with CR.

This is followed, if necessary for that command, by the ASCII parameter "X". If this parameter
"X" is omitted, then the device resets with the current parameter.
A ? after the small command letters answers with the respective settings (only at setting
commands, not at enquiry commands).

Example:

Entry:

“00em“

+ <CR>

The

emissivity

setting

(

e

) of the device with the address 00 is returned.

Answer:

“0970“

+ <CR> means Emissivity = 0.97 or 97.0%

Description

Command

Parameters

Reading
measuring value:

AAms

Output: YYYYY

5-digit decimal, in

1

/

10

°C or. °F

88880 = temp. overflow

02563 = 256.3 °C or °F

Emissivity:

AAemXXXX

XXXX = (0100 ... 1250%) (decimal)

Transmittance:

AAetXXXX

XXXX = (0100 ... 1000%) (decimal)

Exposure time t

90

:

AAezX

X = 0 ... 6 (decimal)
0 = intrinsic time constant of the device
1 = 0.5 s 3 = 2 s 5 = 10 s
2 = 1 s 4 = 5 s 6 = 30 s

Analog output*:

AAasX

X = 0 or 1

0 = 0 … 20 mA;1 = 4 … 20 mA

Clear time of

maximum /
minimum value
storage:

AAlzX

X = 0 ... 8 (decimal)

0 = Maximum value storage off
1 = 0.10 s

4 =

1.00 s

7 = externally deleted

2 = 0.25 s

5 =

5.00 s

8 = automatically deleted

3 = 0.50 s

6 = 25.00 s

External clearing:

AAlx

Delete maximum storage value (only active with lz=7)

Changing °C / °F *: AAfhX

X = 0;1 (decimal)

0 = Output Celsius

1 = Output Fahrenheit