6 controller calibration – LumaSense Technologies M360 User Manual

Page 16

Advertising
background image

M360 Manual

Installation & Operation 16

Each M360 address must be set to a different value. To alter the address if RS485 was ordered:

1. Press and hold PAGE button. ACCESS and Level 1 appear in lower display.
2. Press UP button until LEVEL 3 appears in lower display.
3. Press UP button until 3 appears and wait.
4. Press PAGE until COMMS H appears.
5. Press UP until COMMS J appears. (J COMMS is the RS-485 output)
6. Press SCROLL until ADDRESS appears.
7. Press UP or DOWN to set new address and wait.
8. Press PAGE until ACCESS appears.
9. Press DOWN until LEVEL 1 appears and wait.

Note: A CD containing a sample program is included with your order. For more details

on communications, refer to the electronic communication manual also found on the
CD provided.

3.2.6 Controller Calibration

This blackbody has a precision temperature
sensor monitoring the emitter temperature. The
THERMOCOUPLE measurements are passed
through a look-up table that ultimately
provides the correct exit radiance temperature

for the indicated temperature. A separate
calibration report is included.
This controller has been programmed to contain
a custom look-up table that may be used to
precisely calibrate the controller/indicator
against reference sensor. The look-up table to

the right appears as shown in the controller GUI
interface (not supplied).
The logic of this table is: “when the sensor input
measures the value shown in INPUT #, force the
display to show the value in the corresponding

OUTPUT # and do linear interpolation between
points”. A sample table is shown to the right.
The actual table used for your calibration is
shown on the next page.
Follow directions on the next page to alter
output values from the controller front panel.

Note: You may download Eurotherm iTools Software from this web site:

http://www.eurotherm.com/products/controllers/itools/

With this software and some practice, you will see this GUI interface screen. To permit

iTools to talk to your blackbody, you may have to set the controller comm. port to
“Modbus” mode. Procedure to change communications to Modbus are shown in
Section 3.2.7.