LumaSense Technologies M360 User Manual

Page 17

Advertising
background image

M360 Manual

Installation & Operation 17

Typical Radiance temperature calibration look-up table (LUT) used for model M360

To gain access to the custom look-up table, refer to the 4 square buttons on the face of the
controller and use this procedure.

1. Press and hold 1 until ACCESS appears. If Access Level and Level 1 appears, skip to

step 3.

2. Press 4 until Level 3 appears.
3. Press 4 until 3 appears. Wait until PASS appears.
4. Press 1 until Lin16 appears.
5. Press 2 until display shows the Output # that corresponds to the calibration point shown

in the listing on the previous page.

6. Use UP/DWN buttons (3 and 4) to dial in the measured radiance temperature from a

freezing point transfer device or other radiance standard. Adjust Out# only!

7. Press 2 to view/change more correction points.
8. When finished with all calibrations and to lock the controller, press 1 until ACCESS

appears.

9. Press 3 until Level 1 appears.

Note: Calibration points do not have to be used as pre-programmed. To dial in a

different calibration point, simply change an INPUT # and the corresponding OUTPUT#
to be the new desired calibration point. Set the blackbody to the calibration
temperature using its setpoint. When exit radiance data is measured, dial in the new
OUTPUT # temperature as described above. All cal points must be in ascending order!