M360 Manual

Installation & Operation 18

Example: In place of Silver freezing point correction at 961.78 C, you prefer to use 1000C.

1. Using the above steps, set INPUT 9 equal to 1000.0. Set OUTPUT 9 equal to 1000.0.

2. Allow blackbody to control and stabilize at 1000.0C.

3. With your radiance standard, you measure 1001.2 C.

4. Use the above procedure to dial in 1001.2 in the OUTPUT 9 location.

5. Wait for blackbody to re-stabilize at 1000.0 and verify radiance temperature.

All corrections between calibration points are performed using straight-line interpolation.

3.2.7 Switch Comm Mode from Ei Bisync to Modbus

Place Eurotherm controller in CONFIG mode by pressing:

1. Press and hold button 1 until ACCESS appears.
2. Press 4 until CONFIG appears. Wait.

3. Press 4 until 4 appears. Wait, then ‘pass’ appears.
4. Press 1 until Comms appears.
5. Press 2 to select EIBISYNCH
6. Press 4 until Modbus appears. Wait.
7. Press 1 until ACCESS appears.
8. Press 4 until Level 1 appears. Wait.

To connect to a computer to run iTools:

1. Attach an RS-232 cable to computer using the comm. cable that came with the kit

(straight DB9 cable).

2. Open iTools program on PC.
3. Press SCAN button on upper bar of program.
4. Select YES. If connected properly, you should see hard disk activity when SCAN

channel 1 is on.

5. After the program ‘sees’ the controller (a picture of a controller appears at lower left of

screen).

6. Follow directions in Section 3.2.5 to access the controller radiance temperature LUT

table.

7. Reset the communications mode to EIBISYNCH using the steps above except replace

Modbus with EIBISYNCH in step 6 above.