M360 Manual

Introduction 9

2 Introduction

2.1 Description

The M360 blackbody calibration source uniquely combines portability with wide temperature
range (50° to 1100 °C), high emissivity and remarkable resolution. An optional precision

machined aperture wheel assembly allows different sizes of aperture diameter for applications
requiring specific radiating aperture or for verifying field of view of radiometers or infrared
thermometers. The source and the controller are housed in separate modules which allow the
source to be positioned in a location remote from the controller such as in an environmental
test chamber, or to be used in tests which involve long path lengths. Each module is fitted with
a carrying handle and can be comfortably carried to manufacturing plant or field research

locations.

2.2 Technical Data

Temperature Range:

50 C to 1100 C or 212 F to 2012 F

Accuracy:

±0.2% of reading ±1 °C

Temperature Resolution:

0.1 C

Stability:

1 C per 8 hour period

Source Non-Uniformity:

±0.2 C typical at T<230 C, ±1 °C typical at T>230 °C

Heated Cavity Shape:

Spherical

Aperture Diameter:

[ ]25.4 mm (1.00”) standard

Emissivity (calculated from

cavity shape):

0.995 ±0.0005

Effective emissivity:

1.00 @ 8-14µm T<230 °C, 1.00 @ 0.7-1.8µm T>230 °C

Method of Control:

Digital Self-Tuning PID Controller

Warm-up Time:

40 minutes from ambient to 700 °C

Operating Ambient

Temperature:

0 C to 44 C or 32 F to 110 F

Operating Humidity:

90% RH max, non-condensing

Cooling:

Fan cooled, air inlet on rear panel

Power Requirements:

[ ]120VAC [ ]230/240VAC [ ]60Hz [ ]50Hz

Dimensions:

See drawings in Section 2.3

Weight:

Blackbody: 17.8 kg (39.3 lbs) Controller: 4.86 kg (10.7 lbs)