so, if the Gas Monitor is measuring in the parts-per-billion range, it is
recommended that the external filters are always used to prevent de-
grading of the tubing’s internal surface.
The filter unit is attached to the tubing as follows, see Fig.2.9.
Fig.2.9. Attaching Fitting UD5041 and Filter DS2306 to sampling tubing
1. Push the Fitting UD5041 into the tubing.
2. Screw the short stub of the Filter DS2306 into the Fitting.
2.3.3
Connecting a Tracer Gas Supply
Tracer gases for use with the 1303 must be supplied from a pressurized
cylinder fitted with a suitable pressure-regulator. The regulator ensures
that the tracer-gas pressure is constant.
Warning! Do not attempt to connect a tracer-gas cylinder without a
pressure-regulator directly to the 1303. Doing this can damage the
1303’s internal airways.