1303 Multipoint Sampler and Doser

LumaSense Technologies A/S

Page 55 of 60

Table 4.4 The interface jobs which control the 1303’s temperature-

measurement function.

Job Header

Data

Effect on 1303

Sensor_Temperature? Specific tempera-

ture-sensor number

Reads-out the tem-
perature (°C) at the

specified tempera-
ture-sensor.

4.2.5

1303 Check Jobs

These jobs check the various functions of the 1303. For further infor-
mation, refer to

section 3.8

.

Table 4.5 The interface jobs which check the 1303’s functions.

Job Header

Data

Effect on 1303

Sampling_Pump_Pressure? No data.

Measures and
reads-out the pres-

sure (kPa) across
the sampling pump.

Dosing_Pump_Pressure?

No data.

Measures and

reads-out the pres-
sure (kPa) across

the dosing pump.

Status?

No data.

Reads-out the
1303’s status-flag.

Check_System

No data.

Starts a check of

the 1303’s sampler
system.

Reset_System

No data.

Resets the 1303.

4.2.6

Error Condition and Service Request Jobs

These jobs allow the identification of error conditions in the 1303. For
further information, refer to

section 3.9

.