2. Remove transport securing device

Remove the O-ring which secures the sliding

weights.

Retain this for further transport operations.

4. Place tray on scale

Place the tray centrally on the scale.

The locking levers must engage under the

housing.

6. Release lock

3. Secure sliding weight bar

Secure the sliding weight bar by tilting the

locking device downwards.

5. Attach tray

Attach the tray to the scale by pressing down

the locking levers.

7. Approximate taring

For the scale to display the correct weight, it

must first be set to 0 kg/lbs.

To do so, push both sliding weights on the

sliding weight bar to 0. If the hand and

counter-hand do not line up, the 0 value

must be set using the tare weight. To do so,

undo the knurled nut (see Fig. 8) and push

the tare weight onto the rod until the hand

and counter-hand line up.