Series 72 manual page 17 – COSA Xentaur Hygrocontrol Type 72 User Manual

Page 21

Advertising
background image

Series 72 Manual page 17

Humidity:

Again this value is sent with resolution of 1/10 %RH (or any other dimension)

without decimal point as ASCII-code. The highest decimal is sent first.

Dewpoints may have negative values which are transmitted as their dual-

complement (16-Bit), without sign, in the same way as negative temperatures.

Example:

00809 is to read 80.9 %RH (or other dimension)

9.3. Error Messages via digital Output

If an error occurs while measuring temperature and humidity (Short circuit or

break of line), the corresponding value of digital output will be replaced by

"ERR--" . "--" stands for two spaces, no error number will be sent!

10. Error Messages

10.1. Tests when Transmitter is set to Power

All transmitters of series HYGROCONTROL 72 are supplied with a test routine,

which is activated if the transmitter is powered. Not only are all segments of the

dual line LCD-display switched on for a short period, but also internal and

external memories, and reference resistors - used for temperature

measurements - are checked.

If errors are detected, the program stops any other activity and an error

message is displayed. This message can be one of the list below:

1) "TEST 9"

= Pushbuttons not free = Press buttons, try again.

2) "TEST 8"

= Error of internal RAM

3) "TEST 7"

= Error, reading the internal EEPROM'S

4) "TEST 5"

= Error, reading the external EEPROM'S

5) "TEST 4"

= Reference resistors for Temperature detection are

out of tolerance. (Upper line of LCD-display

shows additional error number!)

6) "TEST 0"

= No error detected = normal display of Humidity-

and Temperature- Values will come shortly!

In addition to message "TEST 9", also message "TEST 5" allows the user to

make an improvement by himself. Please check if a probe is connected to the

transmitter!

10.2. Error Messages in normal Measurement Mode.

During the normal measurement mode, the microprocessor steadily checks all

signals which are necessary for a proper output. If any error occurs, in addition

to the display of the remaining signal (Humidity or Temperature), an error

Advertising