Series 72 Manual page 21

PSF = Saturated Vapor Pressure over Chilled Water
PSE = Saturated Vapor Pressure over Ice
M

= Rel. Humidity displayed by HYGROMESS Instruments.

11.7 Measurements under Pressure or Vaccuum

All probes can be used at pressures of 0,9 to 1,3 bar. Only special probes

with a glas feed-through (which predicts SS - 316 tube material) may be

used under pressures of 0,03 to 30 bar. As rel. Humidity is directly

proportional to the pressure (f.E.: increase of pressure by factor 2 also

increases rel. Humidity by factor 2, when temperature and content of water

remain constant!) you shoud take care of a place with constant pressure

values, when mounting the probe. Be aware that pressure may change,

when diameters of tubes change - and air streams may be turbulent at

these places.

Table 4: Gas Concentrations with Humidity Error < 2.5 %RH

MAXIMUM CONCENTRATION AT

MAK-
CONCENTRATIO
N
ACCORDING TO

continuous load

50 % load

EXPL.-
LIMIT

CHEMICAL

ppm

g/m

3

ppm

g/m

3

ppm

g/m

3

g/m

3

Ethanol (Alkohol)

1000 1,9

3500 6

7000 12

57

Isopropanol

400 0,98

4800 12

10000 25

67

Toluol + Xylol

100 0,38

1300 5 3000 12

53

Gasoline (pure)

300 1,1-1,4

150

200

51

Gasoline super (Automobil)

100

150

32

Ethylenglycol

100 0,26

1200 3 1200 3

80

Acetone

1000 2,4

3300 8

6500 16

56

Ethylacetat

400 1,4

4000

15

8000 30

79

Acetic Acid

10 0,03

800 2 1200 3

107

Ammoniumhydroxid

25

0,02

5500

4

11000

8 -

Chlorine Acid (HCl)

5

0,01

300

0,5

500

0,75 -

Sulfurhydrogen (H

2

S)

10

0,01

350

0,5

700

1 -

Sulfurdioxid (SO

2

)

5

0,01

5

0,01

5

0,01 -

1) MAK = max. Concentration at Labor (at 1 bar and 20 °C)

2) ppm = g/m

3

x 24,04/MOL

MOL = Molweight of Material