Press the SELECT button and read the moisture content on the meter scale. The meter

displays the %MC for two seconds.

To add a reading to the previously stored readings, release the SELECT button if you are in

triggered mode. If you are in continuous mode, the meter will not store a new reading until
the SELECT button is pressed and released again during the reading cycle.

Note:

Readings between 0.1% and 0.5% indicate a sufficiently dry moisture level.
Readings between 0.5% and 1% indicate a borderline situation. Readings
greater than 1% indicate material that is too wet for painting or wallpaper. A
reading above 1% also indicates enough moisture present to allow mold growth
to occur but only if other factors are also present such as a high RH level and
food source.