Detcon TP-424C User Manual

Page 6

Advertising
background image

Ammonia

500 = 0

Diesel Fuel

5000= 0

Demethyl Sulfide

4.4 = 0

Ethylene

200 = 0

Freon 12

1000 = 0

Hydrogen

1000 = 8

Methyl Mercaptan

5 = 8

Sulfur Dioxide

300 = 0

Toluene

32 = 0

Ethanol

500 = 2

3.5 S

PECIFICATIONS

Method of Detection

Solid state metal oxide. Diffusion/adsorption

Electrical Classification

Class I; Div 1, Groups B, C, D

Response Time

50% <45 seconds, 80% <90 seconds

Clearing Time

50% <45 seconds, 80% <90 seconds

Repeatability

± 2ppm or ±10% of Reading (Greater Of)

Range of Sensitivity

PPB to % by volume

Operating Temperature

-40° to +175° F

Accuracy

± 2ppm or ±10% of Reading (Greater Of)

Sensor Life

10 year conditional warranty

Field Wiring

3 conductor; max 10 ohms single conductor resistance

Input Power

22.5-28 VDC

Power Consumption

57 mA normal operation (1.37 watts @ 24 VDC); 81 mA max. (2 watts @ 24 VDC)

3.6 I

NSTALLATION

NOTE: The Detcon MOS sensor can be damaged to the point of not functioning if the power is turned off in
the presence of high levels of humidity for periods exceeding 8 hours. Always protect the sensor with the
Detcon sealing cap and fresh dessicant packet while the sensr is off power. This will avoid permanent sen-
sor damage and help preserve the span calibration.

Optimum performance of ambient air/gas sensor devices is directly relative to proper application installa-
tion and adherence to recommended maintenance schedules including periodic recalibration.

Field Wiring Table
Detcon Model TP-424C solid state H2S sensor assemblies require three conductor connection between
power supplies and electronic controllers. Wiring designators are + (DC), (DC) , and mA (sensor signal).
Maximum single conductor resistance between sensor and controller is 10 ohms. Maximum wire size for
termination in the sensor assembly terminal board is 14 gauge.

AWG

Meters

Feet

20

240

800

18

360

1200

Detcon Model TP-424C Hydrogen Sulfide Sensor PG.6